| Ora Test Srl è anche su: |
|
|
|

V4
Hipot Electrical Safety Analyzer
Everything You Need In a 4-in-1 Electrical
Safety Analyzer for about
half the price you
would expect to pay
Performance
& Economy
The
new Vitrek V4 Hipot Electrical Safety
Analyzer combines AC & DC Hipot with high
current ground bond and insulation
resistance testing, all in one feature rich,
fully automated, easy to use instrument.
The uncompromising
performance of the V4 delivers fast,
consistent, reliable results – test after
test. And it does it from a single test
connection. The V4’s crystal clear LCD user
interface is remarkably easy to use and has
all the information you need to get the job
done right.
Features/Benefits
5kVAC
40mA, 6kVDC 10mA
-
Ground
Bond 3 to 40 Amps
-
Multi-mode IR 50 to 1000V
-
SFI™
Safety Fault Interrupt
-
RS232
Standard, available GPIB
-
High
Speed Discharge
-
36
Position Memory
-
Selectable Arc Detection
-
PLC
Remote Interface
Top of Page
Specifications
AC Hipot
Voltage Range: |
100
to 5000VAC RMS |
Voltage Setting
Resolution: |
5V/step |
Voltage Regulation: |
1% of
reading ± 5V |
Voltage Accuracy: |
1% of
reading ± 10V |
Voltage Monitoring: |
At
Output Terminals |
Output Current: |
30mA
to 40mA 180S max, 10 to 30mA
continuous. Below 500V 10mA max |
Min/Max Current
Limits: |
0.1
to 40mA |
Current Accuracy: |
1% of
reading ± 50mA |
Top of Page
DC Hipot
Voltage Range: |
100
to 6000VDC |
Voltage Setting |
Resolution: 5V/step |
Voltage Regulation: |
1% of
reading ± 5V |
Voltage Accuracy: |
1% of
reading ± 10V |
Voltage Monitoring: |
At
Output Terminals |
Max Output Current: |
10mA
continuous, below 500VDC 2mA
continuous |
Min/Max Current
Limits: |
0.01
to 10mA |
Current Accuracy: |
1% of
reading ± 50mA |
Insulation Resistance
DC Output Voltage: |
50,
100, 500 or 1000V |
Resistance Range: |
1 to
9999mW |
Resistance Accuracy: (% of reading) |
|
50,
100 VDC |
1-50MW 5%
51-200MW 10%
201-1999MW 20% |
500,
1000 VDC |
1-500MW 5%
501-2000MW 10%
2001-9999MW 20% |
Multi-Mode IR™: |
Select 1 of 3 modes - Test passes on
good reading, test fails on bad
reading, pass/fail determined at end
of test time. |
Top of Page
Ground Bond/Continuity
Test Current: |
3 to
40 Amps RMS |
Terminal
Configuration: |
4
wire Kelvin |
Resistance Range: |
0 to
600mW |
Resistance
Resolution: |
100mW |
Resistance
Accuracy: |
1%
reading ± 3mW |
Compliance Voltage: |
8V at
terminals |
Current Resolution: |
0.01A |
Voltage Resolution: |
0.01V |
Top of Page
General
Test Database: |
6
test sequences with up to 6 steps
per test sequence |
Front Panel
Lockout: |
Prevents database edit while in
lockout mode |
Arc Detect: |
10
level trip select |
Remote Interfaces: |
Fully
interactive RS232 - standard. PLC
Interface with Outputs of Testing,
Pass, Fail and Inputs of Start,
Reset - standard. Talk/Listen GPIB
interface - Option GP-V4. |
Timer Accuracy: |
0.25%
± 50mS |
Power Source: |
100,
120, 220, 230VAC (±10% 50/60Hz),
350W max |
Operating
Environment: |
Indoor use, altitude up to 6500 ft
(2000m). Installation Category II.
Pollution Degree 2. |
Operating
Temperature: |
0°C
to 40°C, 80% RH |
Storage
Temperature: |
-10°C
to 70°C, 70% RH |
Dimension: |
17.6"L x 13"W x 5.9"H, (446mmL x
330mmW x 149mmH) |
Weight: |
33lbs
(15kg) net, 40 lbs (18 kg) ship |
Accessories: |
NIST
Cal Cert, TL-20-50 & K-Test Leads,
Instruction Manual and Power Cord. |
Accuracy: |
Stated specifications apply for 1
year at 23°C ± 5°C. |
Calibration: |
ANSI/NCSL
Z540 Certificate of Calibration with
Data provided, traceable to NIST. |
Compliance: |
CE
Mark Certified to EN61010 and EMC
Directives EN50081-1 & EN50082-2. |
Warranty: |
3
Year Extended Warranty with annual
factory calibration. 1 year parts
and labor standard. |
QuickTest™ Software
Option QT-V4: |
Bundled Test Automation Software and
PC Serial Interface Cable. Easy to
load graphical test sequence
development and operating
environment. Provides time/date,
operator, test parameters, test
results and readings. Results may
be viewed, printed out or exported
to a database program. |
|